Hanna HI-9813-6 Portable pH/EC/TDS/Temperature Meter

Hanna HI-9813-6 Portable pH/EC/TDS/Temperature Meter

The HI9813-6 portable meter features a large LCD which displays either pH, EC, TDS or temperature readings along with tutorial instructions. The pH readings are displayed with a 0.1 resolution and an accuracy of +/-0.1 pH while the EC and TDS readings are displayed with a 0.01 mS/cm and 1 ppm (mg/L) resolution and 2% full-scale accuracy. The EC range of the HI9813-6 is from 0.00 to 4.00 mS/cm and TDS is from 0 to 1999 ppm.

Product Description

Hanna HI-9813-6 Portable pH/EC/TDS/Temperature Meter with CAL Check™

Karya Mandiri Techindo has a wide range of Hanna products – search by type, or model, or filter by brand to find the one you are after. If you still can’t find it contact us..

The HI9813-6 portable meter features a large LCD which displays either pH, EC, TDS or temperature readings along with tutorial instructions. The pH readings are displayed with a 0.1 resolution and an accuracy of +/-0.1 pH while the EC and TDS readings are displayed with a 0.01 mS/cm and 1 ppm (mg/L) resolution and 2% full-scale accuracy. The EC range of the HI9813-6 is from 0.00 to 4.00 mS/cm and TDS is from 0 to 1999 ppm. The temperature correction coefficient (β) is fixed at 2 %/oC and allows for automatic temperature compensated measurements of EC and TDS. The HI9813-6 is a basic pH/EC/TDS meter that is calibrated manually to a single point with the use of two trimmers. pH is calibrated to pH 7.01 while EC/TDS is calibrated to either 1.41 mS/cm (1413 µS/cm) or 1500 ppm. The LCD screen has battery life indicator as well as on-screen tutorial messages. The supplied HI1285-6 probe is a polypropylene body, amplified pH electrode with a built-in EC/TDS and temperature sensors. The amplifier for the pH electrode prevents interference from humidity and electrical noise from common sources including from motors, ballasts or pumps. The HI9813-6 is a versatile meter for the agriculture, greenhouse and hydroponics industries.

The HI9813-6 meter also offers the CAL Check feature that alerts the user to a problem with a pH electrode. This feature is important for customers that calibrate only to pH 7.0; if there is a fracture on the pH glass of the electrode, the pH meter will always display pH 7.0 regardless of the solution being measured. This can be disastrous for the person that calibrates at pH 7.0 and takes readings of samples with an expected pH of 7.0. The user will never be aware that there is a problem with the electrode. This can result in inaccurate pH measurements, which for a greenhouse can affect the growth of the plants. Placing the HI1285-6 pH/EC electrode in the HI50021 CAL Check solution helps users determine if the probe needs to be calibrated, cleaned or replaced. When the probe is placed in solution the user presses the “Check” key. The meter runs CAL Check diagnostics and will display either “Probe is OK” or “Clean Probe and Calibrate”. If the reading is around pH 4.0 when the probe is placed in the solution then the probe is broken and needs to be replaced.

hi1285-6 electrode

HI1285-6 polypropylene body amplified pH electrode with built-in EC/TDS and temperature sensors

The HI9813-6 is supplied with a polypropylene body pH/EC/TDS/temperature probe. The pH, EC, TDS, and temperature sensor are housed in a single body that connects to the meter with a DIN connector.

Amplified pH electrode

The pH electrode circuit has a built-in amplifier that  will reduce the effects of electrical noise on the high impedance pH measurement. Examples of sources of electrical noise include motors, ballasts, and pumps which are common in greenhouses.

Amperometric EC/TDS sensor

The EC/TDS readings are performed by an amperometric sensor. An alternating voltage is applied to the sensor and the amount of current that passes between the two stainless steel pins is dependent upon the amount of salts (fertilizer) present. A greater amount of salt present results in an increase in conductance.

Polypropylene body

The polypropylene body houses all the sensors in a single body design and is durable. The probe is gel filled for maintenance free operation. It does not have to be refilled periodically.

  • 3 sensors in a single probe
  • Pre-amplified pH electrode for resistance to electrical noise
  • Gel filled maintenance free pH electrode

Features at-a-glance

  • Water Resistant
  • CAL Check™
  • Variable EC to TDS conversion factor
    – Factor automatically adjusts from 0.56 to 0.78 based on actual EC readings
    – Factor based on 442 curve for natural water
  • Automatic Temperature Compensation
    – All readings are compensated for variations in temperature
  • Low Battery Indicator

Specifications Hanna HI-9813-6 Portable pH/EC/TDS/Temperature Meter

 
SKU HI9813-6
Product Name Portable pH/EC/TDS/Temperature Meter with CAL Check™ – HI9813-6
Quote Required Yes
pH Range 0.0 to 14.0 pH
pH Resolution 0.01 pH
pH Accuracy ±0.1 pH
pH Calibration manual, 1-point through offset trimmer
EC Range 0.00 to 4.00 mS/cm
EC Resolution 0.01 mS/cm
EC Accuracy ±2% F.S.
EC Calibration manual, 1-point through slope trimmer
TDS Range 0 to 1999 ppm (mg/L)
TDS Resolution 1 ppm (mg/L)
TDS Accuracy ±2% F.S.
TDS Calibration manual, 1-point through slope trimmer
EC/TDS Temperature Compensation automatic from 0 to 50°C (32 to 122°F)
Temperature Correction Coefficient β=2%/°C typical
EC to TDS Conversion Factor variable based on reading from 0.56 to 0.72 according to the TDS 442 curve for natural water
Temperature Range 0.0 to 60.0°C / 32.0 to 140.0°F
Temperature Resolution 0.1°C
Temperature Accuracy ±0.5°C /±1°F
Multi-parameter Probe HI 1285-6 combination, amplified pH/EC/TDS/temperature probe with Cal-Check feature, built-in temperature sensor, 8-pin DIN connector and 1 m (3.3′) cable
Battery Type/Life 9V / approximately 150 hours of continuous use
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 145 x 80 x 36 mm (5.7 x 3.1 x 1.4”)
Weight 230 g (8.1 oz)
Ordering Information HI9813-6 is supplied with HI1285-6 pH/EC/TDS/°C probe with Cal-CheckTM, HI70007 pH 7.01 buffer solution sachet (1), HI70442 1500 ppm calibration solution sachet (1), HI70031 1413 µS/cm calibration solution sachet (1), HI700661 electrode cleaning solution sachets (2), HI50021 calibration check solution sachets (2), 9V battery, instructions and rugged carrying case.

Portable pH/EC/TDS/Temperature Meter with CAL Check™ - HI9813-6

 

 


Hanna HI-9813-6 – Manual


Produk asli Kami distributor langsung Hanna Harga murah dan berkualitas bergaransi tentunya. dapat di kirim ke seluruh wilayah indonesia.

Dalam pengiriman produk yang pelanggan beli, sebelumnya sudah kami cek untuk memastikan produk dalam keadaan baik dan siap kirim.

 

Additional Information

Brand

Hanna

Model

HI-9813-6

Made In

Romania

Weight

2 kg

Product Enquiry

TESTIMONI PELANGGAN

alhamdulillah pesanan sudah sampai di Bengkulu. Pengalaman order di karyamandiritechindo.com Respon cepat. Pengiriman Cepat dan Free, Aman & Terpercaya.

Ibu Natalia Dan Ria
PT. AGricinal

Terimakasih karyamandiritechindo.com, orderannya sudah sampai dengan selamat Kurir Syariftama Express sangat ramah

PT. Bumi Bangun Perkasa
Ibu Rizeki

Order yg kedua, seperti biasa produk sangat memuaskan dan pelayannya ramah, Terimakasih Pak Aji Amanda Semoga di orderan berikutnya tetap TOP

PT. Daruna

Puas banget belanja di karyamandiritechindo.com. pengiriman cepat kualitas produk mantap.

PT. Niaga Andalan Serayat

BUAT PERMINTAAN HARGA