Features Lutron EM-9000 Environment Tester
- Environment instrument, multi-function, all in one
- Type K/J thermometer, Humidity/Temp. meter, Anemometer, Light meter, Sound level meter ( optional adapter )
- Thermocouple Thermometer can default to accept type K or type J Temp. probe
- Temperature measurement can default the display unit to ℃ or ℉
- Anemometer can default the display unit to m/S, FPM, Km/h, mph, knot
- Light meter can default the display unit to LUX or Ft-cd
- Meter can default auto power off or manual power off
- 0.1 degree resolution for temperature measurement
- Microcomputer circuit provides intelligent function and high accuracy
- LCD with two display, easy readout
- Manual and auto data logger, with flexible sampling time selection, can save max. 1,600 reading data
- Records Maximum and Minimum readings with recall
- Data hold function for freezing the desired value
- Build in the input socket for DC 9V power adapter
- RS232 PC serial interface
- Few panel buttons, easy operation
- Built-in low battery indicator
- Heavy duty & compact housing case
- Available for the HVAC applications
Specifications Lutron EM-9000 Environment Tester
General Specifications |
---|
Display | LCD size : 44 mm x 29 mm dual function LCD display |
---|---|
Measurement | * Type K/J thermometer * Humidity/Temp. meter * Anemometer with Temp. * Light meter * Sound level meter ( optional adapter ) |
Sampling Time | Manual : Push the data logger button once will save data one time Auto : 1, 2, 5, 10, 30, 60, 600, 1800 3600 seconds. |
Data Logger number | Max. 1,600-point Data logger |
Weight
Dimension |
306 g/0.67 LB. * Include probe and batteries
Meter : 135 x 60 x 33 mm. |
Accessories included | Instruction manual Probe ( EM-910P ) |
Optional accessories | * Batere kotak 9 Volt * Type K thermocouple probe. Accessories TP-01, TP-02A. TP-03, TP-04 * AC to DC 9V adapter. * USB cable, USB-01 * Data Acquisition software, SW-U801-WIN. * Data Logger software, SW-DL2005. * Sound adapter, SL-417 |
Electrical Specifications | |||
Type K/J Thermometer | |||
Sensor Type | Resolution | Range | Accuracy |
Type K
|
0.1 ℃
|
-50.0 to 1300.0 ℃ | ± ( 0.4 % + 0.8 ) |
-50.1 to -199.9 ℃ | ± ( 0.4 % + 1 ) | ||
0.1 ℉
|
-58.0 to 2372.0 ℉ | ± ( 0.4 % + 1.5 ) | |
-58.1 to -327.8 ℉ | ± ( 0.4 % + 1.8 ) | ||
Type J
|
0.1 ℃ –
|
-50.0 to 1100.0 ℃ | ± ( 0.4 % + 0.8 ) |
-50.1 to -199.9 ℃ | ± ( 0.4 % + 1 ) | ||
0.1 ℉
|
-58.0 to 2012.0 ℉ | ± ( 0.4 % + 1.5 ) | |
-58.1 to -327.8 ℉ | ± ( 0.4 % + 1.8 ) |
Anemometer |
A. Air velocity | |||
Measureent | Range | Resolutiontion | Accuracy |
m/S | 0.4 – 25.0 m/s 0.1 | m/s | ± (2% + 0.2 m/s) |
km/h | 1.4 – 90.0 km/h | 0.1 km/h | ± (2% + 0.8 km/h) |
mph ) | 0.9 – 55.9 | mile/h 0.1 | mile/h ± (2% + 0.4 mile/h |
knot | 0.8 – 48.6 | knots 0.1 knots | ± (2% + 0.4 knots) |
FPM | 80 – 4930 ft/min | 1 ft/min | ± (2%+40 ft/min.) |
Temperature | |
Measuring Range | 0 to 50 /32 to 122 ℃ ℃℉ ℉ |
Resolution | 0.1 /0.1 ℃ ℉ |
Accuracy | ± 0.8 /1.5 ℃ ℉ |
Humidity/Temp. meter |
Humidity | |
Measuring Range | 0 % to 95 % R.H. |
Resolution | 0.1 % R.H. |
Accuracy | 70% RH ≧ ± (3% reading + 1% RH). < 70% RH ± 3% RH. |
Temperature | |
Measuring Range | 0 ℃to 50 ℃ /32 ℉ to 122 ℉ |
Resolution | 0.1 ℃ /0.1 ℉ |
Accuracy ± 0.8 /1.5 ℃ ℉ | ± 0.8 ℃ /1.5 ℉ |
Light meter |
Measuring Range
|
LUX | 0 to 20,000 LUX. |
Ft-cd | 0 to 2,000 Ft-cd | |
Resolution
|
LUX | 1 LUX |
Ft-cd | 0.1 Ft-cd | |
Accuracy | ± ( 5% rdg ± 8 dgt ) |